D.C. Municipal Regulations (Last Updated: September 13, 2017) |
Title 22. HEALTH |
SubTilte 22-B. PUBLIC HEALTH AND MEDICINE |
Section 22-B10802. RECOMMENDATION FOR QUALITY ASSURANCE PROGRAMS IN DIAGNOSTIC RADIOLOGY FACILITIES
-
10802.1Quality assurance programs are recommended for all diagnostic radiology facilities.
10802.2A quality assurance program should contain the elements listed in § 10802.3 through 10802.23. The extent to which each element of the quality assurance program is implemented should be determined by an analysis of the facility's objectives and resources conducted by its qualified staff or by qualified outside consultants. The extent of implementation should be determined on the basis of whether the expected benefits in radiation exposure reduction, improved image quality, or financial savings will compensate for the resources required for the program.
10802.3Responsibility and authority for the overall quality assurance program as well as for monitoring, evaluation, and corrective measures should be specified and recorded in a quality assurance manual.
10802.4The owner or practitioner in charge of the facility has primary responsibility for implementing and maintaining the quality assurance program.
10802.5Staff technologists will generally be delegated a basic quality assurance role by the practitioner in charge. Responsibility for specific quality control monitoring and maintenance techniques or quality administration procedures may be assigned, provided that the staff technologists are qualified by training or experience for these duties. The staff technologists should also be responsible for identifying problems or potential problems requiring actions beyond the level of their training. They should bring these problems to the attention of the practitioner in charge, or his or her representative, so that assistance in solving the problems may be obtained from inside or outside the facility.
10802.6In facilities where they are available, physicists, supervisory technologists, or quality control technologists should have a major role in the quality assurance program. Such specialized personnel may be assigned responsibility for day-to-day administration of the program, may carry out monitoring duties beyond the level of training of the staff technologist or, if desired by the facility, may relieve the staff technologists of some or all of their basic monitoring duties. Staff service engineers may also be assigned responsibility for certain preventive or corrective maintenance actions.
10802.7Responsibility for certain quality control techniques and corrective measures may be assigned to personnel qualified by training or experience, such as consultants or industrial representatives, from outside of the facility, provided there is a written agreement clearly specifying these services.
10802.8In large facilities, responsibility for long-range planning of quality assurance goals and activities should be assigned to a quality assurance committee as described in § 10802.22.
10802.9Before purchasing new equipment, the staff of the diagnostic radiology facility should determine the desired performance specifications for the equipment. Initially, these specifications may be stated in terms of the desired performance of the equipment, or prospective vendors may be informed solely of the functions the equipment should be able to perform and asked to provide the performance specifications of items from their equipment line that can perform these functions. In either case, the responses of the prospective vendors should serve as the basis for negotiations to establish the final purchase specifications, taking into account the state of the art and balancing the need for the specified performance levels with the cost of the equipment to meet them. The final purchase specifications should be in writing and should include performance specifications. The availability of experienced service personnel should also be taken into consideration in making the final purchase decisions. Any understandings with respect to service personnel should be incorporated into the purchase specifications. After the equipment is installed, the facility should conduct a testing program, as defined in its purchase specifications, to ensure that the equipment meets the agreed upon specifications, including applicable Federal and State specifications and the records of the acceptance testing should be retained throughout the life of the equipment for comparison with monitoring results in order to assess continued acceptability of performance.
10802.10A routine quality control monitoring and maintenance system incorporating state-of-the-art procedures should be established and conducted on a regular schedule. The purpose of monitoring is to permit evaluation of the performance of the facility's x-ray system(s) in terms of the standards for image quality established by the facility (as described in § 10802.17) and compliance with applicable Federal and State regulatory requirements. The maintenance program should include corrective maintenance to eliminate problems revealed by monitoring or other means before they have a serious deleterious impact on patient care. To the extent permitted by the training of the facility staff, the maintenance program should also include preventive maintenance, which could prevent unexpected breakdowns of equipment and disruption of departmental routine.
10802.11The parameters to be monitored in a facility should be determined by that facility on the basis of an analysis of expected benefits and cost. Such factors as the size and resources of the facility, the type of examinations conducted, and the quality assurance problems that have occurred in that or similar facilities should be taken into account in establishing the monitoring system. The monitoring frequency should also be based upon need and can be different for different parameters.
10802.12Although the parameters to be monitored will vary somewhat from facility to facility, every diagnostic radiology facility should consider monitoring the following five (5) key components of the x-ray system:
(a) Film processing;
(b) Basic performance characteristics of the x-ray unit;
(c) Cassettes and grids;
(d) View boxes; and
(e) Darkroom.
10802.13Examples of parameters of the above-named components and of more specialized equipment that may be monitored are as follows:
(a) For film processing:
(1) An index of speed;
(2) An index of contrast;
(3) Base plus fog;
(4) Solution temperatures; and
(5) Film artifact identification;
(b) For basic performance characteristics of the x-ray unit:
(1) For fluoroscopic x-ray units:
(A) Table-top exposure rates;
(B) Centering alignment;
(C)Collimation;
(D)kVp accuracy and reproducibility;
(E)mA accuracy and reproducibility;
(F)Exposure time accuracy and reproducibility;
(G)Reproducibility of x-ray output;
(H)Focal spot size consistency;
(I) Half-value layer; and
(J)Representative entrance skin exposures;
(2)For image-intensified systems:
(A)Resolution;
(B)Focusing;
(C)Distortion;
(D)Glare;
(E)Low contrast performance; and
(F)Physical alignment of camera and collimating lens;
(3)For radiographic x-ray units:
(A) Reproducibility of x-ray output;
(B) Linearity and reproducibility of mA stations;
(C) Reproducibility and accuracy of timer stations;
(D) Reproducibility and accuracy of kVp stations;
(E) Accuracy of source-to-film distance indicators;
(F) Light or x-ray field congruence;
(G) Half-value layer;
(H) Focal spot size consistency; and
(I) Representative entrance skin exposures;
(4)For automatic exposure control devices:
(A)Reproducibility;
(B)kVp compensation;
(C) Field sensitivity matching;
(D) Minimum response time; and
(E) Backup timer verification;
(c) For cassettes and grids:
(1) For cassettes:
(A) Film or screen contact;
(B) Screen condition;
(C) Light leaks; and
(D) Artifact identification;
(2) For grids:
(A) Alignment and focal distance; and
(B) Artifact identification;
(d) For view boxes:
(1) Consistency of light output with time;
(2) Consistency of light output from one (1) box to another; and
(3) View box surface conditions;
(e) For darkrooms:
(1) Darkroom integrity; and
(2) Safe light conditions;
(f) For specialized equipment:
(1) For tomographic systems:
(A) Accuracy of depth and cut indicator;
(B) Thickness of cut plane;
(C) Exposure angle;
(D) Completeness of tomographic motion;
(E) Flatness of tomographic field;
(F)Resolution;
(G)Continuity of exposure;
(H)Flatness of cassette; and
(I) Representative entrance skin exposures; and
(2) For computerized tomography:
(A) Precision (noise);
(B) Contrast scale;
(C) High and low contrast resolution;
(D) Alignment; and
(E) Representative entrance skin exposures.
10802.14The maintenance program should include both preventive and corrective aspects.
10802.15Preventive maintenance should be performed on a regularly scheduled basis with the goal of preventing breakdowns due to equipment failing without warning signs detectable by monitoring. Such actions have been found cost effective if responsibility is assigned to facility staff members. Possible preventive maintenance procedures are visual inspection of the mechanical and electrical characteristics of the x-ray system (covering such things as checking conditions of cables, watching the tomographic unit for smoothness of motion, assuring cleanliness with respect to spilling of contaminants in the examination room or the darkroom, and listening for unusual noises in the moving parts of the system), following the manufacturer's recommended procedures for cleaning and maintenance of the equipment, and regular inspection and replacement of switches and parts that routinely wear out or fail. The procedures included would depend upon the background of the staff members available. Obviously, a large facility with its own service engineers can do more than an individual practitioner's office.
10802.16For maximum effectiveness, the quality assurance program should make provisions, as described in § 10802.18 for ascertaining whether potential problems are developing. If potential or actual problems are detected, corrective maintenance should be carried out to eliminate them before they cause a major impact on patient care.
10802.17Standards of acceptable image quality should be established. Ideally, these should be objective. Acceptability limits for the variations of parameter values, but they may be subjective (for example, the opinions of professional personnel, in cases where adequate objective standards cannot be defined). These standards should be routinely reviewed and redefined as needed, as described in § 10802.23 of this chapter.
10802.18The facility's quality assurance program should include means for two (2) levels of evaluation:
(a)On the first level, the results of the monitoring procedures should be used to evaluate the performance of the x-ray system(s) to determine whether corrective actions are needed to adjust the equipment so that the image quality consistently meets the standards for image quality. This evaluation should include analysis of trends in the monitoring data as well as the use of the data to determine the need for corrective actions on a day-by-day basis. Comparison of monitoring data with the purchase specifications and acceptance testing results for the equipment in question is also useful; and
(b)On the second level, the facility quality assurance program should also include means for evaluating the effectiveness of the program itself. Possible means include ongoing studies of the retake rate and the causes of the repeated radiographs, examination of equipment repair and replacement costs, subjective evaluation of the radiographs being produced, occurrence and reasons for complaints by radiologists, and analysis of trends in the results of monitoring procedures such as sensitometric studies. Of these, ongoing studies of the retake rate (reject rate) and its causes are often the most useful and may also provide information of value in the first level of evaluation. Such studies can be used to evaluate potential for improvement, to make corrections, and to determine whether the corrective actions were effective. The number of rejects should be recorded daily or weekly, depending on the facility's analysis of its needs. Ideally, the reasons for the rejection should also be determined and recorded. Should determining these reasons be impossible on a regular basis with the available staff, the analysis should be done for a two (2)-week period after major changes have occurred in diagnostic procedures or the x-ray system and at least semi-annually.
10802.19The program should include provisions for the keeping of records on the results of the monitoring techniques, any difficulties detected, the corrective measures applied to these difficulties, and the effectiveness of these measures. The extent and form of these records should be determined by the facility on the basis of its needs. The facility should view these records as a tool for maintaining an effective quality assurance program and not view the data in them as an end in itself but rather as a beginning. For example, the records should be made available to vendors to help them provide better service. More importantly, the data should be the basis for the evaluation and the reviews suggested in §§ 10802.18 and 10802.23.
10802.20A quality assurance manual should be written in a format permitting convenient revision as needed and should be made readily available to all personnel. The content of the manual should be determined by the facility staff, but the following items are suggested as providing essential information:
(a)A list of the individuals responsible for monitoring and maintenance techniques;
(b)A list of the parameters to be monitored and the frequency of monitoring;
(c)A description of the standards, criteria of quality, or limits of acceptability that have been established for each of the parameters monitored;
(d)A brief description of the procedures to be used for monitoring each parameter;
(e)A description of procedures to be followed when difficulties are detected to call these difficulties to the attention of those responsible for correcting them;
(f)A list of the publications in which detailed instructions for monitoring and maintenance procedures can be found. Copies of these publications should also be readily available to the entire staff, but they should be separate from the manual. (Publications providing these instructions can usually be obtained from the Department or private sources, although the facility may wish to make some modifications to meet its needs more effectively);
(g)A list of the records, with sample forms, that the facility staff has decided should be kept. The facility staff should also determine and note in the manual the length of time each type of record should be kept before discarding; and
(h)A copy of each set of purchase specifications developed for new equipment and the results of the acceptance testing for that equipment.
10802.21The program should include provisions for appropriate training for all personnel with quality assurance responsibilities. This should include both training provided before the quality assurance responsibilities are assumed and continuing education to keep the personnel up-to-date. Practical experience with the techniques conducted under the supervision of experienced instructors, either in the facility or in a special program, is the most desirable type of training. The use of self-teaching materials can be an adequate substitute for supervised instruction, especially in continuing education programs, if supervised instruction is not available.
10802.22A facility whose size would make it impractical for all staff members to meet for planning purposes should consider the establishment of a quality assurance committee whose primary function would be to maintain lines of communication among all groups with quality assurance or image production or interpretation responsibilities. For maximum communication, all departments of the facility with x-ray equipment should be represented. The committee may also be assigned policy-making duties such as some or all of the following:
(a)Assign quality assurance responsibilities;
(b) Maintain acceptable standards of quality; and
(c)Periodically review program effectiveness. Alternatively, the duties of this committee could be assigned to an already-existing committee such as the Radiation Safety Committee. In smaller facilities, all staff members should participate in the committee's tasks. The Quality Assurance Committee should report directly to the head of the radiology department, or, in facilities where more than one (1) department operates x-ray equipment, to the chief medical officer of the facility. The committee should meet on a regular basis.
10802.23The facility's quality assurance program should be reviewed by the Quality Assurance Committee or the practitioner in charge to determine whether its effectiveness could be improved. Items suggested for inclusion in the review include:
(a) The reports of the monitoring and maintenance techniques to ensure that they are being performed on schedule and effectively. These reports should be reviewed at least quarterly;
(b)The monitoring and maintenance techniques and their schedules to ensure that they continue to be appropriate and in step with the latest developments in quality assurance. They should be made current at least annually;
(c)The standards for image quality to ensure that they are consistent with the state-of-the-art and the needs and resources of the facility. These standards should be evaluated at least annually;
(d)The results of the evaluations of the effectiveness of the quality assurance actions to determine whether changes need to be made. This determination should be made at least annually; and
(e) The quality assurance manual should also be reviewed at least annually to determine whether revisions are needed.