Section 22-B10724. DEVICE MASTER RECORD
-
10724.1Each manufacturer shall maintain device master records (DMR). Each manufacturer shall ensure that each DMR is prepared and approved in accordance with § 10706. The DMR for each type of device shall include, or refer to the location of, the following information:
(a) Device specifications including appropriate drawings, composition, formulation, component specifications, and software specifications;
(b) Production process specifications including the appropriate equipment specifications, production methods, production procedures, and production environment specifications;
(c) Quality assurance procedures and specifications including acceptance criteria and the quality assurance equipment to be used;
(d) Packaging and labeling specifications, including methods and processes used; and
(e)Installation, maintenance, and servicing procedures and methods.