Section 22-B10725. DEVICE HISTORY RECORD  


Latest version.
  •  

    10725.1Each manufacturer shall maintain Device History Records (DHRs). Each manufacturer shall establish and maintain procedures to ensure that DHRs for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the DMR and the requirements of this chapter. The DHR shall include, or refer to the location of, the following information:
     

    (a) The dates of manufacture;

     

    (b) The quantity manufactured;

     

    (c) The quantity released for distribution;

     

    (d) The acceptance records which demonstrate the device is manufactured in accordance with the DMR;

     

    (e) The primary identification label and labeling used for each production unit; and

     

    (f)Any device identification(s) and control number(s) used.

     

authority

Section 19(a)(3) of the District of Columbia Pharmacist and Pharmacy Regulation Act of 1980, effective September 16, 1980 (D.C. Law 3-98; D.C. Official Code § 47-2885.18(a)(3) (2005 Repl.)); Mayor’s Order 98-48, dated April 15, 1998; Section 4902 of the Fiscal Year 2002 Budget Support Act of 2001, effective October 3, 2001 (D.C. Law 14-28; D.C. Official Code § 7-731 (2008 Repl.)); Section 15 of the District of Columbia Medical Device Manufacture and Distribution Licensure Act of 1990, effective June 13, 1990 (D.C. Law 8-137; D.C. Official Code § 48-714(a) (2005 Repl.)); and Mayor’s Order 98-88, dated May 29, 1998.